Criticare 8100EP1 nGenuity™ Vital Signs Monitor with Capnography
The 8100EP1 nGenuity Series patient monitors are the perfect choice for busy offices demanding accurate, reliable performance.
The monitor displays true end-tidal CO2 with numerical ETCO2 and INCO2 values as well as a capnographic waveform. A smart respiration feature automatically defaults to the best available source, either CO2 or ECG TTI.
Criticare's advanced design capnometer module is accurate with breath rates up to 120 breaths per minute. Flow rate is set at 200 mL per minute for exceptionally fast response. Rise time responsiveness is within 170 milliseconds, providing outstanding waveform display.
The capnometer modules use side stream sampling and Criticare's patented WaterChek™, which traps water to remove moisture without distortion to the CO2 waveform. The water traps have a large 22 mL reservoir and full occlusion clearing capability unlike many other CO2 monitoring systems.
Compact and lightweight, the nGenuity 8100E1 Series patient monitor can track adult and pediatric patient vital signs in virtually every office and operatory. The innovative user interface features simplified menus and dedicated function keys for timely set-ups. The big, bright display with large numeric's can easily be seen, even from across the room.
The nGenuity mounts quickly to walls and roll stands.
Included Accessories:
- 3 cuffs
- Water trap-sample lines
- 2 sets of ecg pads
- Charger
- Finger sensor-probe
- Nasal cannula
Custom Configurations:
- Standard: ECG, ComfortCuff™ NIBP, DOX™ SpO2, HR, and printer
- Optional: ST/arrhythmia and temperature
Manufactured by Criticare Systems, Inc.
ECG
- Connectors: 3- or 5-lead
- Leads: User-selectable; I, II, III, aVR, aVL, aVF, V (5-lead); I, II, or III (3-lead)
- Gain Selection: x1/2, x1, x2, and x4
- Electrosurgery Protection: Yes
- Electrosurgery Protection: Yes
- Defibrillator Protection: Yes
- Pacer Detection/Rejection: Yes
Non-Invasive Blood Pressure
- Technique: Oscillometric measure upon inflation
- Automatic Measurement Cycles: 2, 3, 5, 10, 15, 30 min; 1, 2, 4 hr
- Measurement Range: Adult: 30 to 300mmHg; Neonate: 0 to 150 mmHg
- Resolution: 1mmHg
- Transducer Accuracy: ±2mmHg or ±2%, which ever is greater
- STAT mode: 5 min of consecutive readings
Pulse Oximetry
- Range: 1% to 99%
- Resolution: 1%
- Accuracy: 70% to 99% range: ±2%; 50% to 69% range: ±3%; <50%: unspecified
- Method: Dual wavelength LED
Heart (Pulse) Rate
- Source: Smart, ECG, Pleth, NIBP
- Range: 20 to 300bpm (30-240bpm, NIBP)
- Accuracy: ±1bpm or ±1% (ECG); ±2 bpm or ±2% (NIBP), which ever is greater
Temperature
- Channels: 1
- Range, Accuracy: 68° to 113° F (20° to 45° C): ±0.1° C
- Probe: YSI-400 and YSI-700 series
Respiration
- Rate: 6 to 150 breaths/min
- Resolution: 1 breath/min Accuracy: ±1% or ±1 breath/min, whichever is greater
- Source: Smart, CO2, ECG-TTI
Capnometry (CO2)
- Units: mmHg, %, kPa, Torr
- Method: Non-dispersive infrared; autocalibrating
- N2O Compensation: Manual; On/Off; user selectable
- Range: 0 to 99 mmHg; 0 to 12.5%; 0 to 12.5 kPa; 0 to 99 Torr
- Resolution: 1 mmHg; 0.1%; 0.1 kPa, 0.1 Torr
- Flow, Sample Rate: Sidestream; 200 ml/min
- Accuracy: ±2 mmHg, ±0.3 volume %, ±0.3 kPa, or ±2 Torr, or ±4% of reading, whichever is greater, at 200 ml/min & respiratory rate <= 120 bpm
- Rise Time: 170 ms (10-90%) at 200 ml/min
Alarms
- Indication: Audible; visual
- Silence: Yes; 2 min or permanent
Trend Memory
- Tabular: 24 hr at 30 sec intervals
- Display: Tabular, Graphical
Display
- Screen: 10.4" active color TFT
- Resolution: 640 x 480 pixels
- Waveforms: 6, maximum
- Languages: English, French, German, Italian, Portuguese, Spanish, Russian, Chinese
Controls
- Keys: 9; membrane-activated
- Rotary Knob: Push and rotate; 24 steps/turn
System Outputs
- Com Ports: RS 232-compatible; DB9 (Com 1); Mini-DIN 8 (Com 2)
- Nurse Call: Contact switch; 3.5mm jack, standard 14mm length
- Defib/Sync: BNC connector
- Video Port: High density D-sub 15-pin
Printer
- Type: Thermal line printer
- Paper Speed: 12.5, 25, 50mm/sec
Mechanical
- Weight: 14 lb; 6.4kg
- Size: Base unit: 11" (H) x 13" (W) x 10.3" (D); 28cm (H) x 33cm (W) x 26.2cm (D)
Power Requirements
- Voltage: 100, 120, 220, 240 VAC; 50/60 Hz
- Consumption: 35 W, typical
- Batteries: 1 sealed lead acid; life: 3 hr, typical recharge time: 4.5 hr
Environmental
- Operating Temperature 59° to 95° F; 15° to 35° C
- Storage Temperature: 23° to 122° F; -5° to 50° C
- Operating Humidity: 15% to 90%, noncondensing
Classification
- Medical Device: Class II Equipment (IIb EU)
- Electrical Protection: Class I Equipment
- Degree of Protection: Type CF defib-proof
- Protection Against Ingress: IPX1
Specifications subject to change without notice.
Customers who bought this also bought...
|
Criticare Finger Sensor, Pediatric
$150.00
|
GE PROCARE B40 Vital Signs Monitor
$0.00
|
Safe Sedate® N2O Nasal Masks, Large (case of 12)
$95.00
|
King Vision Kit
$1,599.00
|






